Abstract

We performed a search for monopolar electrocautery and magnetic resonance imaging safety recommendations for several indwelling neuromodulatory devices. We developed a questionnaire followed by an educational compendium and a posttest for 50 care providers. Overall, there was a poor performance on the pretest (mean 39%, SD 19%) but significant improvement on the posttest (mean 71%, SD 16%), P < 0.0001. We placed the educational compendium that included all manufacturer recommendations in the operating room for easy reference. A 2.4 times decrease in the case start times of vagus nerve stimulator cases is evidence of its effectiveness. The authors highlight the lack of knowledge about manufacturer safety recommendations for indwelling neurosurgical devices, which led to the creation of operating room supplements and educational devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.