Abstract

Microplastic pollution has become an environmental problem that cannot be ignored in our society. Raman spectroscopy technology has been widely used in the field of microplastics detection due to its non-contact, non-destructive chemical specificity. Traditional point confocal Raman micro-spectroscopy technology uses single-point detection, resulting in long measurement times to scan the large areas of interest of typical samples. In this paper, we present a line scan confocal Raman micro-spectroscopy tool for fast detection and identification of microplastic particles. We show size and composition identification of particles and imaging over large areas. Compared with point confocal Raman imaging, the line scan confocal Raman technology increases the imaging speed by 1–2 orders of magnitude.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.