Abstract

A system for rapid, high precision potentiometric determination of alkalinity in sea water and sediment pore water is presented. Two titration units were used: a 40 ml unit for seawater and a small volume unit for sediment pore water. Titration time was normally less than 10 minutes per sample, including sample exchange. With a 40 ml sample volume, the relative standard deviation of the alkalinity obtained in the laboratory was 0.05% and at sea 0.1 %. The small-volume system (0.5–1.5 ml) gave a precision of 0.07%. Five titration points, in two groups after the second equivalence point, were used to evaluate the equivalence volume. Results from equilibrium calculations and computer simulated alkalinity titrations show that it was possible to use a non-modified Gran function [( V 0 +v) ∗10 ( E Z ) ] and still achieve good accuracy and precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.