Abstract
A simple and sensitive procedure was developed for extraction and speciation of mercury in fish. Species separation was accomplished with reverse phase-high performance liquid chromatography (HPLC) hyphenated to inductively coupled plasma mass spectrometry (ICP-MS). Oxygen addition into plasma allowed use of organic-rich mobile phase, achieving species separation in 4min. Mercury species extraction was achieved by microwave exposure for 2min at mild conditions (60°C, pH 2.0), avoiding necessity of neutralizing sample prior to injection in HPLC, and reducing number of sample preparation steps, analytical source of errors and inter conversion of species. Limit of detection for entire procedure was found to be 0.2 and 0.1ngg−1 for mercuric ion and methylmercury, respectively. The method was applied to certified reference materials (TORT-2 and DORM-2) and commercialized fish samples (Mullus barbatus, Sparus aurata, Trachurus mediterraneus, Mugil soiuy, Dicentrarchus labrax, and Pomatomus saltatrix) from Black Sea.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.