Abstract

A new, simple, and rapid chiral HPLC method was developed for enantioselective analysis of levetiracetam and its enantiomer [(R)-α-ethyl-2- oxo-pyrrolidine acetamide] in a pharmaceutical formulation and bulk material. Enantiomeric separation was achieved on a chiral-α1-acid glycoprotein (AGP) column (150×4.0 mm, 5 μm) using an isocratic mobile phase of phosphate buffer (pH=7) at a flow rate of 0.7 mL/min. The UV detector was set at 210 nm. Calibration curves were linear in the range of 1-100 μg/mL and 0.4-20 μg/mL for levetiracetam and the (R)-enantiomer, respectively. LOD and LOQ for the (R)-enantiomer were 0.1 and 0.4 μg/mL, respectively. The run time of analysis was less than 5.0 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.