Abstract

In this study we describe a novel method of sampling and quantifying wound biomarkers for clinical settings. We believe the chosen format will allow rapid assessments of wound healing and provide biomarker evidence-based decision points for treatment of the wound at the time of presentation. The wound monitoring principle uses a proprietary sample collection tool (a thermally reversible hydrogel) to sample and isolate biomarkers within a wound environment without further sample extraction/preparation steps. We show how gel samples can be analysed in a lateral flow assay format utilising fluorescent microspheres with optically discrete emission characteristics and demonstrate quantitative detection of two analytes (duplexing) achieved in a single test line. As a model assay, the chronic wound biomarkers interleukin 6 (IL6) and tumour necrosis factor alpha (TNFα) are used. Limits of detection of 48.5pg/mL and 55.5pg/mL respectively in hydrogel samples and 7.15pg/mL and 10.7pg/mL respectively in plasma are reported. We believe this is the first literature example of quantitative detection of multiple analytes within a single test line using spectral separation to distinguish the analytes.