Abstract

BackgroundUltraviolet germicidal irradiation (UVGI) technologies have emerged as a promising adjunct to manual cleaning, however, their potential to shorten cleaning times remains unexplored. MethodsA <10-minute disinfection procedure was developed using a robotic UVGI platform. The efficacy and time to perform the UVGI procedure in a CT scan treatment room was compared with current protocols involving manual disinfection using biocides. For each intervention, environmental samples were taken at 12 locations in the room before and after disinfection on seven distinct occasions. ResultsThe mean UVC dose at each sample location was found to be 13.01 ± 4.36 mJ/cm2, which exceeded published UVC thresholds for achieving log reductions of many common pathogens. Significant reductions in microbial burden were measured after both UVGI (P≤.001) and manual cleaning (P≤.05) conditions, with the UVGI procedure revealing the largest effect size (r = 0.603). DiscussionThese results support the hypothesis that automated deployments of UVGI technology can lead to germicidal performance that is comparable with, and potentially better than, current manual cleaning practices. ConclusionsOur findings provide early evidence that the incorporation of automated UVGI procedures into cleaning workflow could reduce turnaround times in radiology, and potentially other hospital settings.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call