Abstract

A rapid technique is presented for measuring organic C and N in carbonate-bearing sediments. Organic C and N were determined with a Yanaco MT-5 CHN analyzer after removing carbonate by acidification of samples in ceramic sample boats with HCl vapor. This procedure eliminates loss of acid-soluble organic matter during carbonate dissolution. The accuracy of this new technique and two earlier methods of determining organic C and N concentrations was verified with standard chemicals and carbonate-bearing sediments (> 80 wt% CaCO3). With the new technique there was no contamination or loss of organic C and N. The method is suitable for acidifying sediments with a wide range of carbonate content.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.