Abstract

Direct and rapid analysis of cholesterol was accomplished in the ambient environment using reactive desorption electrospray ionization (DESI) mass spectrometry. This was achieved by electrospraying reagent solutions in the form of high velocity charged droplets at surfaces such as dried serum samples and animal tissue sections. Betaine aldehyde, incorporated into the spray solvent, reacts selectively and rapidly with the alcohol group of cholesterol by nucleophilic addition, forming a hemiacetal salt. Limits of detection for pure cholesterol and related compounds were approximately 1 ng when a solution of cholesterol of 1 microg/mL was spotted onto the surface. Quantitative analysis of free cholesterol in serum using reactive DESI was demonstrated using cholesterol-d7 as internal standard. High throughput analysis of small volumes of serum spotted onto a suitable substrate was achieved at an analysis rate of approximately 14 s per sample, with a relative standard deviation (RSD) of ca. 6%. Use of reactive DESI in the imaging mode allowed 2D spatial distributions of phospholipids and cholesterol to be recorded simultaneously in rat brain tissues.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.