Abstract

Tiaprofenic acid (Surgam) has been determined directly by using differential pulse polarography (DPP) and pH 5.5 acetate buffer as a supporting electrolyte. The differential pulse polarogram was obtained under constant amplitude pulses of 50 mV superimposed on a linearly increasing dc ramp. Peak current was measured at the peak potential of -0.990 V vs Ag/AgCl reference electrode. Under the experimental conditions used, a linear relationship between peak current and concentration was established over the range 0.5-5.0 micrograms/mL. Mean percentage recoveries for tiaprofenic acid in authentic and tablet forms were 100.27 +/- 2.25 and 99.45 +/- 1.44, respectively. The results obtained by the DPP method were compared with those of the spectrophotometric method used by the manufacturer for the analysis of tiaprofenic acid and its tablet form.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.