Abstract
The intrinsic heights equivalent to a theoretical plate (HETPs) of 31 narrow-bore and wide-bore columns packed with four different brands of core–shell particles were accurately measured on an optimized vHPLC instrument (1290 Infinity system) that has an extra-column volume variance of 13.6±0.3μL2. These results were derived from the slopes of the linear plots of the apparent plate heights of each column versus the reciprocal of (1+k′)2 for seven homologous compounds with a linear regression coefficient larger than 0.999. The results show that the kinetic performance of narrow-bore columns packed with core–shell particles increases almost linearly with decreasing particle diameter. The optimum reduced plate heights increase slightly from 1.6 to 1.9 with decreasing particle sizes from 4.6 to 1.3μm. This confirms that wide-bore columns provide better efficiencies than narrow-bore columns.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.