Abstract

This paper describes a rapid, simple and sensitive analytical method for the quantitative determination of melatonin in human serum by ESI–MS–MS with direct serum sample injection and on-line extraction. The method uses N-acetyltryptamine as the internal standard. It has high specificity and sensitivity for serum melatonin analysis. The internal calibration curve shows a wide linear range from 0.500 to 200 ng/ml with a correlation coefficient, R 2>0.999. The limit of quantitation is 0.500 ng/ml and the limit of detection is 0.100 ng/ml with 10-μl sample injection. The recoveries of serum melatonin at three levels are approximately 70%. The intra-assay precision ( n=5) is between 0.8 and 2.0% and the inter-assay precision ( n=3) is between 1.5 and 5.9% over the calibration range. This method has a total analysis time of less than 9 min. It can be used for the measurement of melatonin in human blood.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.