Abstract

High resolution inductively coupled plasma mass spectrometry coupled with an on-line flow injection system (FI-HR-ICP-MS) was applied to determine the ultra-trace level 99Tc in soil. The flow injection system (PrepLab™) was composed of two TEVA-SpecOresins, reduced remarkably the sample amounts and the analysis time, compared to the conventional analytical methods. In the flow injection system, Mo and Ru were sufficiently eliminated by using the flow injection system, with the decontamination factors of 1.6.104 and 9.9.105, respectively. With the present method, it was possible to determine ultra-low level of 99Tc in 3-6 g soil at 3–5 hours of analysis time per sample. The relative standard deviation for each sample was less than 4%. The detection limits for 99Tc was 85 fg.ml–1 (0.05 mBq.ml–1), which was calculated from the three times standard deviation of the count rate of the blank.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.