Abstract

A new method has been developed at the Savannah River National Laboratory (SRNL) that can be used for the rapid determination of 226Ra in emergency urine samples following a radiological incident. If a radiological dispersive device event or a nuclear accident occurs, there will be an urgent need for rapid analyses of radionuclides in urine samples to ensure the safety of the public. Large numbers of urine samples will have to be analyzed very quickly. This new SRNL method was applied to 100 mL urine aliquots, however this method can be applied to smaller or larger sample aliquots as needed. The method was optimized for rapid turnaround times; urine samples may be prepared for counting in 90 %), and removes interferences effectively. The sample preparation method can also be adapted to ICP-MS measurement of 226Ra, with rapid removal of isobaric interferences.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.