Abstract

Rapid photometric assessment of iron in blood plasma and serum by a simple procedure after the extraction of iron(II) complex with 1-nitroso-2-naphthol in the micellar phase of a nonionic surfactant at the cloud point upon heating (pH range is 4.5–6.3) is proposed. The procedure trueness was verified using a standard reference protocol using bathophenanthroline. The advantages of the procedure are higher sensitivity than the reference protocol: the limit of detection is 0.03 μg/mL, the limit of quantitation is 0.1 μg/mL, the determination range is 0.1 – 2.8 μg/mL (RSD 0.02–0.10). Copper does not interfere with the iron assessment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.