Abstract

A novel assay is described for multiplex detection of antibodies against different pathogens from a single sample. The assay employs a modified lateral flow format (consecutive flow, CF) together with a sensitive reporter particle technology (up-converting phosphor technology, UPT) that allows for fully instrumented assay analysis. Lateral flow (LF) strips developed for the detection of human antibodies against human immunodeficiency virus type-1 and -2 (HIV-1 and -2) with additional capture zones to detect antibodies against Myobacterium tuberculosis (TB) and hepatitis C Virus (HCV) provided the strips to test multiplexing. Data are presented that show the performance of the TB and HCV test, as well as two multiplex assays, TB with HIV and HCV with HIV. The TB/HCV assays demonstrate excellent detection capability, and HIV multiplexing does not affect the qualitative test result. The bench-top CF format was converted to a microfluidic platform and a first prototype semiautomated chip capable of performing CF is presented here.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.