Abstract

The design and performance of the high-brilliance XAFS facility at BL10XU of SPring-8, aimed at rapid and sensitive measurement of X-ray absorption fine structure (XAFS), is reported. Both undulator gap and double-crystal monochromator have been successfully controlled covering a wide energy range (5-30 keV). A versatile goniometer system, consisting of two independent high-precision goniometers, is capable of polarized XAFS in fluorescence mode and surface-sensitive experiments using a grazing-incidence geometry. By sharing major components, i.e. a monolithic Ge 100-pixel array detector and a closed-cycle He cryostat, both polarized XAFS and X-ray standing wave (XSW) experiments can be performed at low temperature (15-300 K). The performance of the spectrometer has been evaluated by recording XAFS spectra in transmission mode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.