Abstract

The present study was aimed to develop a loop-mediated isothermal amplification (LAMP) assay for rapid and specific detection of Vibrio cholerae. A set of five designed primers that recognized specifically the V. cholerae ompW gene was used. The optimized time and temperature conditions for the LAMP assay were 75 min at 65 degrees C, respectively. The LAMP method accurately identified 16 isolates of V. cholerae but did not detect 28 non-cholerae Vibrio isolates and 37 non-Vibrio bacterial isolates. The sensitivity of LAMP for V. cholerae detection in pure cultures was 2.2 x 10(3) CFU ml(-1) or equivalent to 8 CFU per reaction. In the case of spiked shrimp samples without enrichment, the detection limit for V. cholerae was 2.2 x 10(4) CFU g(-1) or equivalent to 20 CFU per reaction, while that of PCR was 100 CFU per reaction. The developed LAMP assay targeting ompW gene was rapid, specific and sensitive for V. cholerae detection. The developed LAMP assay appears to be precise, accurate and a valuable tool for detection of V. cholerae. This assay can replace laborious biochemical tests for the identification of V. cholerae in contaminated food sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.