Abstract
A fully automated immunoassay of 17β-estradiol (E2) was performed using anti-E2 monoclonal antibody immobilized on bacterial magnetic particles (AntiE2-BMPs) and alkaline phosphatase-conjugated E2 (ALP-E2). E2 concentration in environmental water samples was evaluated by decrease in luminescence based on competitive reaction. A linear correlation between the luminescence intensity and E2 concentration was obtained between 0.5 and 5 ppb. The minimum detectable concentration of E2 was 20 ppt. All measurement steps were done within 0.5 h. The analysis of environmental water samples by a commercially available ELISA kit and the BMP-based immunoassay gave good correlation plots with a correlation efficient of 0.992. These results suggest that the fully automated system using the BMP-based immunoassay has some advantages in the high rapidity and sensitivity of the measurement. This system will enable us to determine low E2 concentrations without sample condensation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.