Abstract

Carbonate clumped isotope abundance is an important paleothermometer, but measurement is difficult, slow, and subject to cardinal mass (m/z) interferences using isotope ratio mass spectrometry (IRMS). Here, we describe an optical spectroscopic measurement of carbonate clumped isotopes. We have adapted a tunable infrared laser differential absorption spectrometer (TILDAS) system to measure the abundances of four CO2 isotopologues used for clumped isotope thermometry. TILDAS achieves the same precision (0.01‰ SE) as IRMS measurements rapidly (∼50 min per carbonate analysis) and using small samples (<2 mg of calcite), without making assumptions about 17O abundance in the sample. A temperature calibration based on 406 analyses of CO2 produced by digestion of 51 synthetic carbonates equilibrated at 6° to 1100°C is consistent with results for natural carbonates and previous calibrations. Our system results were indistinguishable from IRMS systems after replicating the InterCarb interlaboratory calibration. Measurement by TILDAS could change the landscape for clumped isotope analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.