Abstract

The high precision measurement of doubly substituted ("clumped") isotopologues in CO2 is a topic of significant interest in isotope geochemistry. Here we describe the performance of a new isotope ratio laser spectrometer using tunable infrared laser differential absorption spectroscopy (TILDAS). The TILDAS instrument has two continuous-wave lasers to simultaneously measure the four isotopologues involved in the 12C16O2 + 13C16O18O ⇆ 13C16O2 + 12C16O18O exchange reaction. CO2 samples are trapped in a low volume (∼250 mL) optical multipass cell with a path length of 36 m. Each sample is compared to a reference gas, and clumped isotopologue precision of 0.01‰ (SE) is achieved within 20 min for 15 μmol samples. Similar precision is also achieved for bulk isotopic composition. The degree of rare isotope clumping in excess of strictly random distribution (Δ16O13C18O) measured by this TILDAS instrument varies linearly with theoretically calculated values and shows a very weak dependence on bulk isotopic composition.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.