Abstract

Nine pesticides of commonly used in Gaza Strip were tested in this work for optimal separation onto silica gel TLC plates in acetone/ cyclohexane, 1:2 (v/v), as a mobile phase to develop a new densitometric method for quantification and determination of these pesticides in water samples. These pesticides are: abamectine, atrazine, benomyl, cymoxanil, diuron, imidacloprid, linuron, penconazole, and propazine. Calibration plots were linear between 25 to 1000 ng/ spot for all pesticides with a correlation coefficient, r, between 0.9920 and 0.9993. Recoveries after enrichment of real water samples on Bakerbond RP-C18 cartridges were between 79.4 and 103.7 % at a relative standard deviation between ±1.92 and 3.68 %. Detection limits LOD of 25 - 100 ng per spot for eight of the pesticides were reached, which meet the European limits for pesticides in drinking water (100 ngL-1), while penconazole has an LOD value of 200 ngL-1 because of its low UV-absorption coefficient.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.