Abstract

This paper describes the development and performance of a new rapid amperometric biosensor for fructose monitoring in food analysis. The biosensor is based on the activity of fructose dehydrogenase (FDH) immobilised into a carbon nanotube paste electrode according to two different procedures. The direct wiring of the FDH in a highly original osmium-polymer hydrogel was found to offer a better enzyme entrapment compared to the immobilisation of the enzyme in an albumin hydrogel. The optimised biosensor required only 5U of FDH and kept the 80% of its initial sensitivity after 4months. During this time, the biosensor showed a detection limit for fructose of 1μM, a large linear range between 0.1 and 5mM, a high sensitivity (1.95μAcm−2mM), good reproducibility (RSD=2.1%) and a fast response time (4s).Finally, the biosensor was applied for specific determination of fructose in honey, fruit juices, soft and energy drinks. The results indicated a very good agreement with those obtained with a commercial reference kit. No significant interference was observed with the proposed biosensor.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.