Abstract

In early 2022, the number of people infected with the highly contagious mutant severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), called Omicron, was increasing worldwide. Therefore, several countries approved the lateral flow assay (LFA) strip as a diagnostic method for confirming SARS-CoV-2 instead of reverse transcription-polymerase chain reaction (RT-PCR), which takes a long time to generate the results. However, owing to the limitation of detection sensitivity, commercial LFA strips have high false-negative diagnosis rates for patients with low virus concentrations. Therefore, in this study, we developed a portable surface-enhanced Raman scattering (SERS)-LFA reader based on localized surface plasmon effects to solve the sensitivity problem of the commercial LFA strip. We tested 54 clinical samples using this portable SERS-LFA reader, which generated 49 positive and 5 negative results. Out of the 49 positive results, SERS-LFA classified only 2 as false negative, while the commercial LFA classified 21 as false negative. This confirmed that the false-negative rate had significantly improved compared to that of commercial LFA strips. We believe that the proposed SERS-LFA system can be utilized as a point-of-care diagnostic system to quickly and accurately determine a virus infection that could spread significantly within a short period.