Abstract
A model is presented that predicts analyte concentration rapidly from slowly equilibrating sensor technology. Specifically, the model characterizes response from a thin coating film deposited on a quartz crystal microbalance (QCM) device and applies directly to ion exchange and adsorption phenomena, but can be modified to any slowly equilibrating system. Several cases are evaluated with emphasis on chemical and biochemical analyte detection. Results demonstrate that the model can predict analyte concentrations after step changes within a few percent of the equilibration time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.