Abstract

A simple and rapid method for analysis of malathion in blood was developed using head space-solid phase microextraction (HS-SPME) and gas chromatography mass spectrometry/ electron impact ionization-selected ion monitoring (GC-MS/EI-SIM). A vial containing a blood sample, ammonium sulphate, sulphuric acid and fenitrothion as an internal standard, was heated at 90 °C for 15 min. The extraction fiber of the SPME was exposed for 5 min in the head space of the vial. The compounds absorbed on the fiber were detached by exposing the fibre in the injection port of GC-MS. A straight calibration curve was obtained between malathion concentrations of 2.5 to 50.0 μg g −1 in blood. No interfering substances were found, and the time for analysis was 40 min for one sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.