Abstract

BackgroundOut-of-hospital cardiac arrest is a leading cause of mortality and serious neurological morbidity in Europe. We aim to investigate the effect of 3 cardiopulmonary resuscitation (CPR) feedback devices on effectiveness of chest compression during CPR. MethodsThis was prospective, randomized, crossover, controlled trial. Following a brief didactic session, 140 volunteer nurses inexperienced with feedback CPR devices attempted chest compression on a manikin using 3 CPR feedback devices (TrueCPR, CPR-Ezy, and iCPR) and standard basic life support (BLS) without feedback. ResultsComparison of standard BLS, TrueCPR, CPR-Ezy, and iCPR showed differences in the effectiveness of chest compression (compressions with correct pressure point, correct depth, and sufficient decompression), which are, respectively, 37.5%, 85.6%, 39.5%, and 33.4%; compression depth (44.6 vs 54.5 vs 45.6 vs 39.6 mm); and compression rate (129.4 vs 110.2 vs 101.5 vs 103.5 min−1). ConclusionsDuring the simulated resuscitation scenario, only TrueCPR significantly affected the increased effectiveness compression compared with standard BLS, CPR-Ezy, and iCPR. Further studies are required to confirm the results in clinical practice.