Abstract

SummaryWe compared the performance of the Ambu® AuraOnceTM Laryngeal Mask with that of the LMA ClassicTM laryngeal mask airway during controlled anaesthesia. Forty patients requiring intermittent positive pressure ventilation were studied using a randomised crossover design. The mean (SD) oropharyngeal leak pressure for the Ambu device (19 (7.5) cmH2O) was significantly greater than for the LMA Classic (15 (5.2) cmH2O; p = 0.004), and the number of attempts for successful insertions was significantly less (39 (50%) vs 45 (56%), respectively; p = 0.02). There was one failure to obtain a patent airway with the Ambu Laryngeal Mask and none with the LMA Classic. Insertion of the Ambu Laryngeal Mask required more manipulations to achieve a patent airway than did the LMA Classic (6 (15%) vs 1 (2.5%), respectively; p = 0.045), but the time taken for insertion was similar between the two groups. The incidence of trauma, grade of fibreoptic view, peak airway pressure and quality of ventilation during maintenance of anaesthesia were similar in both groups.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.