Abstract

During the past decade, the use of portable Raman analyzers for field measurements has grown dramatically. However, most analyzers use 785 nm excitation lasers that can cause permanent eye damage. To overcome this safety concern, we have built a portable Fourier transform (FT) Raman analyzer using a 1550 nm retina-safe excitation laser and have compared its performance to our 1064 nm FT-Raman analyzer, which uses the same optical design. Raman theory predicts approximately five times lower peak intensities at 1550 nm. Although we found that intensities were as much as 20 times less intense, the analyzer is still capable of measuring spectra of sufficient quality to identify and differentiate chemicals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.