Abstract
The vertical flow (VF) method improves generation and collection efficiency in Raman spectroscopy. It enhances all Raman signals, including undesired signals of organic solvents having a considerably large Raman cross section. We constructed a Raman spectrometer using the VF method to overcome this drawback and introduced a spatial line rejection mask to eliminate unnecessary bands. In addition, the design of the VF unit was improved to resist organic solvents. A VF unit with a 60-µm pinhole enhanced the signal 168 times. The spatial mask effectively eliminated the large Raman bands of the solvent and enabled a longer exposure time. The increase in the dynamic range improved the signal-to-noise ratio by 10 % in methanol and acetonitrile measurements. Raman spectrometer with the VF method and spatial mask enables us to record the Raman spectrum of solute molecules without the disturbance of solvent bands.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.