Abstract

A high concentration of radon in groundwater has attracted recent attention as a precursor of seismic activity. We have constructed a system that extracts and counts radon gas from solid, liquid, and gas samples. The radon is extracted in a closed system onto activated charcoal. The desorbed radon is then measured in a phosphored acrylic cell by scintillation counting of gross alpha radiation. The efficiency of the total system (extraction plus counting) is 90 +- 3% or better. Compact design and sturdy construction make the system completely portable and well suited to field operations in remote loations. Results are given for radon and radium in groundwaters in the Livermore area.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.