Abstract
We proposed an iterative correlation coefficient matching strategy for the high-precision measurement of the radius of curvature (ROC), which only needs a single-frame interferogram. This strategy includes two parts: one is to obtain an initial ROC by using the relationship between the ROC and the diameter of the dark ring; other is to further solve a high-precision value by an iterative method. In the iterative part, we simulated a series of interference patterns with each ROC in the setting range, and calculated the correlation coefficient between the each simulated interferogram and the measured interferogram, and searched the maximum of the correlation coefficient that would be used to confirm a new ROC in the next iteration. When the calculated ROCs in the adjacent iterations meet the termination condition, the high-accuracy ROC can be solved. The feasibility of our proposed method has been verified by simulations and experiments. The errors of the test mirrors with the ROC of 101.6 mm and 1000 mm are 0.013% and 0.024% in experiments, respectively. This research provides a simple and feasible method for realizing the high-precision ROC measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.