Abstract
Determining the radiochemical purity (RCP) of technetium-99m (99mTc) radiopharmaceuticals using the method described in the package insert is a time-consuming process, requiring particular attention in order to achieve accurate RCP results. The purpose of this study was to evaluate whether radio-ultra high performance liquid chromatography (radio-UHPLC) may be an alternative method for RCP testing of 99mTc-tetrofosmin, 99mTc-MAG3 and 99mTc-sestamibi. Results obtained using radio-UHPLC were in excellent agreement with the standard method, with total analysis time being reduced to less than 3min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.