Abstract

Changes in radiostrontium clearance (SrC) and bone formation (tetracycline labeling) were observed in the femurs of skeletally mature dogs following the various operative steps involved in bone screw fixation. Drilling, but not periosteal stripping, produced a small but statistically significant increase in SrC and endosteal bone formation in the distal third of the bone. Strontium clearance values equivalent to those produced by drilling alone were recorded after screw fixation at low or high torque (5 versus 20 inch pounds), as well as by the insertion of loosely fitting stainless steel implants. Bone formation (equals the percentage tetracycline-labeled trabecular bone surfaces) was increased by 30% when SrC values exceeded 3.5 ml/100 g bone/min, and the relationship was linear when SrC values ranged between 1.0 and 7.0 ml/100 g bone/min. The changes in SrC and bone formation one-week after bone screw application are primarily those associated with a response to local trauma caused by drilling.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.