Abstract

PurposeThe study proposes the use of three-dimensional CdZnTe Compton camera (3D-CZT CC) for radiopharmaceutical imaging and investigates the influence factors using a 3D-printed mouse phantom. MethodsThe event selection method and image reconstruction algorithm are optimized by Monte Carlo simulations and mouse phantom experiments. ResultsSimulation results show that the intrinsic energy resolution and spatial resolution of 3D-CZT cause a certain deviation in the calculated Compton scattering angle and Compton axis. Such deviation causes the imaging quality to deteriorate. By selecting events whose distance between Compton and photoelectronic interactions are larger than 10 mm, the mean deviation of the Compton axis could be reduced to less than 10%. Using the ordered origin ensemble algorithm with resolution recovery, the artifacts around organs where the radiopharmaceutical was placed are reduced, and the quality of the reconstruction results are improved compared to the results with simple back projection and origin ensembles algorithms. The phantom study shows that the 3D-CZT CC imaging device could visualize the radiopharmaceuticals distribution by 15 min detection. ConclusionsThrough the analysis of this study, the feasibility of 3D-CZT CC for in-vivo distribution measurement of radiopharmaceuticals is demonstrated, and the quality of reconstruction result has been improved.