Abstract
Radionuclidic purity tests of 18F radiopharmaceuticals (Na18F and fluorodeoxyglucose [18F]FDG) and radionuclide composition analysis of irradiated water [18O]H2O were performed. The measurements were conducted using a High-Purity Germanium (HPGe) detector and a liquid scintillation counter. Radionuclide identification and activity measurements were performed for samples from different stages of the production process. Most of the impurities were detected on QMA (quaternary methylammonium) anion exchange columns and in liquid wastes. Using liquid scintillation counting, the activity of 3H resulting from the 18O[p, 3H]16O reaction was determined. It was shown that all of the impurities were efficiently determined and eliminated in the radiopharmaceuticals synthesis process and that the final products meet the requirements set by relevant regulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.