Abstract

Portable nonimaging probes have been developed that can evaluate left ventricular function using radionuclide techniques. Two modes of data acquisition are possible with these probe systems, first-pass and gated. Precordial radiocardiograms obtained after a bolus injection can be used to determine cardiac output, pulmonary transit time, pulmonary blood volume, left ventricle ejection fraction, and left-to-right shunts. Gated techniques can be used to determine left ventricular ejection fraction and sytolic time intervals. Probe-determined indices of left ventricular function agree excellently with comparable measurements determined by conventional camera-computer methods as well as by invasive techniques. These have begun to be used in a preliminary manner in a variety of clinical problems associated with left ventricular dysfunction. This review discusses the types of probe systems available, the methods used in positioning them, and details the specifics of their data acquisition and processing capacity. The major criticisms of probe methods are that they are nonimaging and that they measure global rather than regional left ventricular function. In spite of these criticisms, probe systems, because of their portability, high sensitivity, and relatively low cost are useful supplements to conventional camera-computer systems for the measurement of parameters of left ventricular performance using radionuclide techniques.