Abstract

A suspension of 224Ra adsorbed onto CaCO3 microparticles shows promise for α-therapy of intracavitary micro-metastatic diseases. To facilitate accurate activity administrations, geometry-specific calibration factors for commercially available reentrant ionization chambers (ICs) have been developed for 224RaCl2 solutions and 224Ra adsorbed onto CaCO3 microparticles in suspension in ampoules, vials, and syringes. Ampoules and vials give IC responses consistent with each other to <1%. Microparticles attenuation leads to a ≈1% to ≈2.5% reduction in response in the geometries studied.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.