Abstract

When two hyperspectral cameras are sensitive to complementary portions of the electromagnetic spectrum it is fundamental that the calibration processes conducted independently lead to comparable radiance values, especially if the cameras show a shared spectral interval. However, in practice, a perfect matching is hard to obtain, and radiance values that are expected to be similar might differ significantly. In the present study we propose to introduce an ulterior linear correcting factor in the radiometric calibration pipeline of two hyperspectral cameras, operating in the visible near infrared (VNIR) and short wave infrared (SWIR) intervals. The linearity properties of both cameras are preliminarily assessed, conducting acquisitions on five standardized targets, and highlighting noise at the sensors level and different illumination fields as the main causes of radiance mismatch. The correction step that we propose allows the retrieval of accurate and smoothly connected VNIR-SWIR reflectance factor curves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.