This paper describes a radiometric determination of the melting/freezing point of gold by measurement of the spectral radiance of a black body at that temperature. A simple optical system was used to image the black-body aperture onto a filter radiometer. The advantages and disadvantages of this approach are discussed. The techniques to characterize the optical system are described together with the assessment of uncertainties. The freezing point of gold was found to be 1 337,330 K ± 28 mK.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.