Abstract

A rapid vacuum-driven procedure, using pre-treated Sep-Pak C18 cartridges, has been developed for the simultaneous extraction of arginine vasopressin (AVP) and atrial natriuretic peptide (ANP) from plasma. Non-specific interference was removed by fractional elution with an aqueous methanol/trifluoroacetic acid (TFA) mixture. AVP and ANP were coeluted under positive pressure with a methanol/TFA mixture and the eluates air-dried before measurement using separate radioimmunoassays. Assay ranges for AVP and ANP were 0.12-29.5 pmol/L and 0.65-162 pmol/L, respectively, with mean recoveries (standard deviation in parentheses) for AVP of 96.4% (5.5%) at a level of 11.8 pmol/L and for ANP of 94.8% (5.9%) at a level of 32.4 pmol/L. The extraction and assay procedures were validated by observing the changes in plasma AVP and ANP concentrations in normal subjects at different stages of hydration and in elderly patients during treatment for congestive cardiac failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.