Abstract

A reliable radioimmunoassay for the determination of plasma 21-deoxycortisol (21-DF) after chromatography on Sephadex LH20 columns of methylene chloride plasma extracts has been described and evaluated. The antiserum used was raised in rabbits injected with 21-DF-3-(O-carboxy-methyl)oxime-bovine serum albumin. In men (n = 10) the levels ranged from 0.11 to 0.29 ng/ml (mean +/- SD: 0.21 +/- 0.06). In women the mean levels were in the follicular phase: 0.16 +/- 0.06 (range: 0.05-0.22; n = 10) and in the luteal phase: 0.18 +/- 0.06 (range: 0.10-0.35; n = 12). No cyclical change and no significant difference between male and female groups were observed. In patients with congenital adrenal hyperplasia due to 21-hydroxylase deficiency very high levels were observed. The higher concentration of 21-DF found in adrenal effluent than in peripheral plasma has provided direct evidence of its adrenal origin in patients without 21-hydroxylase deficiency.