Abstract

We report the development of a double-antibody system for radioimmunoassay of CK-B subunit in isoenzymes CK-MB and CK-BB of creatine phosphokinase. With our method, 4.1 ng/ml of isoenzyme CK-BB can be detected. Within-run and between-run coefficients of variation are respectively 5.3% and 19.6% Analytical recovery of added antigen is 98.9 ± 14.7%. The CK-B concentration in sera from 20 healthy adults was less than 75 ng/ml. In sera from 25 patients with a rise in CK-total activity, but without CK-MB or CK-BB activity tested with the previous reported immunological methods, we found CK-B concentrations in the range 0–86 ng/ml. In contrast, in 23 sera from patients with acute myocardial infarction within the last 48 hours CK-B concentration was in the range 104–225 ng/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.