Abstract

A radioimmunoassay (RIA) for measurement of platelet-activating factor (PAF) was developed. At a final antiserum dilution of 1:640, the lowest detection limit of PAF was 0.1 pmol (50 pg). The standard curve obtained was suitable for measurement of PAF in amounts ranging from 0.1 pmol to 30 pmol. The antiserum showed high specificity. Cross-reaction for lysoPAF, lysophosphatidylcholine and long-chain phosphatidylcholines was very low (less than 0.025%). 1-Palmitoyl-2-acetyl-sn-glycero-3-phosphocholine cross-reacted slightly (6.25%). PAF exogenously added to macrophage suspensions was quantitatively determined by RIA after solvent extraction and high-performance liquid chromatographic separation. RIA was also used to estimate PAF formation after stimulation of rabbit alveolar macrophages in suspension with calcium ionophore A23187.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.