Abstract

Knee wear is commonly measured with the most recent radiograph based on the assumption that wear progresses at a constant rate. Changing patient activity or in vivo polyethylene deterioration are examples of factors that could cause wear rates to change over time. Using six or more radiographs on each of 251 knees over a mean 10-year follow-up, we determined the pattern of polyethylene wear. 92% of knees had linear wear with a mean wear rate of 0.09 ± 0.12 mm/yr. Ten knees (4%) had late accelerated wear. Knees with accelerated wear and those with a wear rate of 0.15 mm/yr or greater had lower survivorship rates. We conclude that wear is linear with rare exceptions and that higher wear is correlated with failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.