Abstract

PurposeTo measure the luminance level of X-ray viewing boxes and ambient lighting levels in reporting rooms as a quality assurance procedure, and to compare the results with those recommended by the Directorate of Radiation Control, South Africa (DRC), European Commission (EC) and Nordic Radiation Protection Co-operation (NORDIC).Materials and MethodsAll the viewing boxes housed at the Divisions of Radiation Oncology and Radiology of Johannesburg Hospital had their luminance levels measured using a calibrated photometer. In addition the room’s ambient light was measured using a photometer.ResultsThe mean average luminance was 1026.75 ± 548.65 cd m-2 and 3284.38 ± 327.91 cd m-2 at the Division of Radiology and Division of Radiation Oncology respectively. The Division of Radiation Oncology had an average viewing box uniformity of 7.14% compared to 27.32% at the Division of Radiology. The average ambient lighting was found to be 66.30 lux and 66.43 lux at the Division of Radiation Oncology and Division of Radiology respectively.ConclusionThe radiograph viewing conditions in Johannesburg variably comply with guidelines. This study underscores the need to implement quality control and quality assurance standards in radiographic image viewing.