Abstract
It is not generally appreciated that there is a gradation of the focal spot size from the anode to cathode end of an x-ray field that can dramatically affect radiographic detail from one end of a radiograph to the other. The authors name this gradation the "focal effect." Gradations in radiation intensity and focal spot size were measured from anode end to cathode end of a 14 x 17-inch field and were visually demonstrated with lymphangiograms and line-pair patterns. The degradation in spatial resolution along the anode-cathode axis was found to be as much as 75%, depending on the orientation of the patient with respect to the x-ray tube. Radiographic detail is, therefore, significantly improved (even when the large focal spot is used) by exploiting the focal effect and placing the body part requiring the best radiographic detail at the anode end of the table.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Radiographics : a review publication of the Radiological Society of North America, Inc
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.