Abstract

To evaluate and compare the maximum temperature (T(max) ) in the head after exposure to a 300 MHz radiofrequency (RF) field induced by a magnetic resonance imaging (MRI) coil using two thermal simulation methods: Pennes' bioheat equation (PBHE) and discrete vasculature (DIVA). The electromagnetic field induced in the head by a 7T birdcage coil was simulated using finite-difference time-domain (FDTD) and validated by MRI. The specific absorption rate (SAR) distributions normalized to the 10-gram maximum or the whole-head average were used for PBHE and DIVA simulations. For all cases, the T(max) in PBHE was slightly higher than in DIVA. The T(max) was 37.9-38.4°C, depending on the simulation method or perfusion rate. In some situations, RF exposure limited to SAR(max,10g) led to a T(max) higher than allowed by International Electrotechnical Commission (IEC) regulations. Therefore, it is advisable to use thermal simulations to evaluate RF safety of MRI. The simulation method used only slightly influenced the observed maximum temperature; the observed temperature with PBHE was higher in all situations. So PBHE is an appropriate method for RF safety assessment of MRI in the head. Using DIVA simulations, it was found unlikely that the body temperature increases significantly due to energy deposited by a head coil under normal circumstances.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.