Abstract

The dose distribution produced by the high-dose-rate (HDR) 192Ir source is inherently anisotropic due to self-absorption by the high-density source core, oblique filtration by the source capsule and the asymmetric geometry of the source capsule. To account for the dose distribution anisotropy of brachytherapy sources, AAPM Task Group No 43 has included a two-dimensional anisotropy function, F(r, θ), in the recommended dose calculation formalism. Gafchromic HS radiochromic film (RCF) was used to measure anisotropy function for microSelectron HDR 192Ir source (classic/old design). Measurements were carried out in a water phantom using specially fabricated PMMA cylinders at radial distances 1, 2, 3, 4 and 5 cm. The data so generated are comparable to both experimental and Monte Carlo calculated values for this source reported earlier by other authors. The RCF method described in this paper is comparatively high resolution, simple to use and is a general method, which can be applied for other brachytherapy sources as well.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.