Abstract

Very low levels of plutonium ( 239,240Pu and 238Pu) in marine samples (sea water,sediments, marine organisms) are determined by extraction on columns of tri-n-octyl phosphine oxide supported on microporous polyethylene (Micro thene-710), electrode- position, and α-spectrometry. 236Pu or 242Pu is added as the yield detector and a high-resolution α-spectrometer is used for counting. The final recoveries are 62.6% ± 9.7(σ)for sea water, 45.4% ± 9.6(0) for sediments and 81.7%± 4.5(σ) for marine organisms. Themethod enables 239.240pu and 238Pu to be detected at the femtocurie level.